Quality

Our quality system is built to ISO/IEC 17025, and our application for ISO/IEC 17025 accreditation is in progress. This page states how we handle impartiality, confidentiality, metrological traceability, measurement uncertainty, decision rules, proficiency testing, competence, records and complaints.

ISO/IEC 17025: application in progress

Our application for ISO/IEC 17025 accreditation is in progress.

Accreditation has not yet been granted. No test described on this site is inside a scope of accreditation, and our reports carry no accreditation mark.

ISO/IEC 17025:2017, General requirements for the competence of testing and calibration laboratories, is the international standard used to assess testing laboratories. An accreditation body assesses a laboratory's quality system and its technical competence for a defined list of tests, the scope of accreditation, before it grants accreditation.

Our quality system is built to the standard. The stages are: quality system built to the standard (done), application for accreditation (in progress), assessment (next), accreditation decision (after the assessment).

When accreditation is granted, the accredited scope will be listed on this page.

Fig. 1Status of our application for ISO/IEC 17025 accreditation. The solid step is complete, the hatched step is where we are, and the dotted steps are still ahead.
1Quality system built to the standardDone2Application for accreditationIn progress3AssessmentNext4Accreditation decisionAfter the assessment1 Quality system built to the standardDone2 Application for accreditationIn progress3 AssessmentNext4 Accreditation decisionAfter the assessment

Impartiality

Risks to impartiality are identified, recorded and managed under our quality system, and staff declare any conflict of interest.

Pricing never depends on results.

ISO/IEC 17025 clause 4.1 requires a laboratory to identify risks to its impartiality on an ongoing basis and to show how it removes or reduces them.

Confidentiality

Client data are confidential and are released only to the client or as the client directs.

Where the law requires us to release information, we tell the client what was released, unless the law prohibits it.

Metrological traceability

Metrological traceability means that a result can be related to a stated reference through a documented, unbroken chain of calibrations (ISO/IEC 17025 clause 6.5).

Quantitative results are linked to the SI through calibrated balances and volumetric ware, and to a characterized reference standard. Where no certified reference material exists, as for erinacine A, the report names the reference standard and its certificate.

Reference standards are held with their certificates of analysis, logged on receipt, stored and used within their expiry.

A plate count is defined by its method and has no chemical reference standard. Its measured inputs are the mass of the test portion, the dilution volumes and the incubation temperature, and the media and methods are checked with reference strains.

Measurement uncertainty

Every quantitative result has a measurement uncertainty: an interval around the result that is expected to contain the true value at a stated level of coverage.

The method of evaluation is described in the Guide to the expression of uncertainty in measurement (JCGM 100:2008) and in the Eurachem/CITAC guide Quantifying Uncertainty in Analytical Measurement (third edition, 2012). Each contribution is expressed as a standard uncertainty; for the erinacine A example these are reference standard purity, calibration, repeatability, loss on drying, volumetric steps and weighing. The contributions are combined as the root sum of squares, and the combined value is multiplied by a coverage factor k = 2, which gives about 95 % coverage.

In the worked example, the erinacine A result is 20.0 mg/g on a dry-weight basis with an expanded uncertainty of ±0.8 mg/g (k = 2). The example is a simulated run with illustrative data, not client data.

For plate counts, ISO 19036:2019 describes the estimation of uncertainty on log10-transformed counts.

A detection result, reported as detected or not detected, is given without a numerical uncertainty. Its reliability rests on the verification of the method and on the positive and negative controls run with the batch.

The method used and the limit of quantitation are stated on the report. Measurement uncertainty is reported on request, and whenever it affects a statement of conformity, as an expanded uncertainty with its coverage factor.

Fig. 2The result is on a dry-weight basis. Relative standard uncertainties combine as the root sum of squares: uc = 1.977 %. Expanded with k = 2 (about 95 % coverage), U = 2 × 1.977 % = 3.95 %. The bars show the size of each u, not its share of the variance. Illustrative data.
Reference standard purity1.00Calibration1.20Repeatability1.10Loss on drying0.40Volumetric steps0.30Weighing0.10Combined1.977Expanded, k = 23.9501234Relative standard uncertainty (%)Reference standard purity1.00Calibration1.20Repeatability1.10Loss on drying0.40Volumetric steps0.30Weighing0.10Combined1.977Expanded, k = 23.9501234Relative standard uncertainty (%)
Uncertainty budget, erinacine A
Reference standard purity1.00 %
Calibration1.20 %
Repeatability1.10 %
Loss on drying0.40 %
Volumetric steps0.30 %
Weighing0.10 %
Combined, uc, root sum of squares1.977 %
Expanded, U = 2 × uc3.95 %

Erinacine A, reported20.0mg/g, dry weightExpanded uncertainty U = ±0.8 mg/g (k = 2)

3.95 % of 20.0 mg/g = 0.79 mg/g, reported as ±0.8 mg/g

Decision rules and statements of conformity

A statement of conformity says whether a result meets a specification. Because every result has an uncertainty, a result close to a limit needs a rule that says how the uncertainty is taken into account. This is the decision rule.

When a client asks for a statement of conformity, ISO/IEC 17025 requires the specification and the decision rule to be defined and agreed with the client when the request is reviewed, before testing (clause 7.1.3). The report then identifies the results the statement applies to, the specification, and the decision rule applied (clause 7.8.6). ILAC-G8:09/2019, Guidelines on Decision Rules and Statements of Conformity, describes the rules in common use.

Unless a different rule is agreed, a result conforms only when its whole uncertainty interval (k = 2) lies inside the specification. The acceptance limit therefore sits one expanded uncertainty, U, inside the specification limit. For a single specification limit, the probability of stating conformity for an item that does not conform is then no more than 2.5 %.

A result closer to the limit than U is not stated as conforming. The report gives the result and its uncertainty.

A client can agree a different rule before testing, such as simple acceptance, in which the result is compared directly with the limit. For a result exactly at the limit, the risk of a wrong decision is then 50 %. Where the rule is part of the specification or standard itself, as in a microbiological criterion with a sampling plan, that rule applies.

We report against the specification or limit the client supplies, or against a named reference at the client's request. We do not advise which limit applies to a product. Results, and any statement of conformity, relate only to the items tested.

Fig. 3The decision rule. Unless a different rule is agreed, a result conforms only when its whole uncertainty interval (k = 2) lies inside the specification. For a specification of at least 18.0 mg/g with U = 0.8 mg/g, that means a result of at least 18.8 mg/g. The rule is agreed before testing and named on the report. The specification shown is an example, not a recommended limit. Illustrative data.
Specification: at least 18.0 mg/gAcceptance limit 18.8 mg/g (guard band = U)Outside the specification20.0 ± 0.8 mg/g (k = 2)19.220.81718192021Erinacine A (mg/g)Specification:at least 18.0 mg/gAcceptance limit 18.8(guard band = U)Outside thespecification20.0 ± 0.8 mg/g (k = 2)19.220.81718192021Erinacine A (mg/g)

Proficiency testing and internal quality control

ISO/IEC 17025 clause 7.7 requires a laboratory to monitor the validity of its results inside the laboratory and, where schemes are available and appropriate, by comparison with other laboratories.

In a proficiency-testing round a provider sends portions of the same material to many laboratories and scores each result against the assigned value.

We are enrolling in proficiency-testing schemes and will judge each result by its z-score. Under ISO 13528:2022, a z-score with an absolute value of 2.0 or less is acceptable, a value between 2.0 and 3.0 is a warning signal, and a value of 3.0 or more is an action signal.

Where no scheme covers an analyte, as for erinacines, we arrange interlaboratory comparisons.

Inside the laboratory, every check standard is plotted on a control chart. A trend is visible on the chart before a control fails its criterion.

Fig. 4Check-standard recovery over 30 consecutive sequences. Center line 100.0 %; limits from the method’s historical data (s = 1.1 %): warning ±2s at 97.8 / 102.2 % (dashed), action ±3s at 96.7 / 103.3 % (solid). In this example a check standard outside 95.0–105.0 % fails the run; a point beyond the action limits stops the method for investigation even when it is inside acceptance. Illustrative data.
Action 103.3 %Warning 102.2 %Center 100.0 %Warning 97.8 %Action 96.7 %Point 17: beyond the warning limit, reviewed,within the action limitthis sequence96979899100101102103104Recovery (%)151015202530SequenceActionWarningCenterWarningActionPoint 17: beyond the warninglimit, reviewed, within theaction limitthis sequence9698100102104Recovery (%)151015202530Sequence

Competence of analysts

Each analyst is trained and then authorized method by method, and competence is checked again with blind samples and replicate work.

A second analyst reviews every result before it is signed.

Documents and records

Methods, procedures and forms are versioned, reviewed and approved before use. Superseded versions are withdrawn and archived.

Every entry records who made it and when, the first capture of the data is kept, and every correction is traceable, so that a reported result can be followed back to the original observation. All data are kept, including repeat injections and failed runs.

Records are stored on durable, backed-up media for the retention period and can be retrieved for review or audit.

Available on request: quality manual extract, method summaries, validation summaries and the impartiality policy.

Complaints, nonconforming work and review

Every complaint is logged, investigated by someone not involved in the work, and answered in writing.

Nonconforming work is traced to its root cause, corrected, and checked for effect.

The whole system is audited on a planned schedule, and management reviews the results, complaints and corrective actions at least once a year.

A complaint can be sent through the contact details on the Contact page.