Methods and instruments

This page describes the techniques used across the six testing categories, the controls run with every batch, one UPLC-PDA determination followed step by step, and how data are recorded and reviewed. Every result comes from a documented method, is calibrated inside its range where calibration applies, and is checked by a second analyst before it is signed.

Techniques

UPLC-PDA

Ultra-performance liquid chromatography with photodiode-array detection

Instrument
Waters ACQUITY UPLC H-Class Bio with PDA detector
How it works

A clarified extract of the sample is injected into a stream of solvent and pumped at high pressure through a column packed with particles smaller than 2 µm. The compounds interact with the packing to different degrees and leave the column one after another; in gradient elution the solvent is made steadily stronger during the run. The photodiode-array detector records an absorbance spectrum at every point of the run, so each peak has a retention time, a spectrum and an area. The area at one wavelength is converted to a concentration with a calibration prepared from a reference standard. For mycotoxins, the extract is cleaned up before injection and the detection mode is stated on the report.

What controls the result

Before any sample is injected, system suitability injections show that resolution, peak shape and injection repeatability meet the criteria of the method. A peak is identified only when both its retention time and its UV spectrum match the reference standard. Calibration is external, and the sample injections are bracketed by check standards; a sample that reads above the calibrated range is diluted and injected again, never extrapolated.

GC-FID/MS

Gas chromatography with flame-ionization and mass-spectrometric detection, with headspace sampling

Instrument
Gas chromatograph with headspace sampler, flame-ionization detector and mass spectrometer
How it works

The sample is vaporized and carried by an inert gas through a heated capillary column, where compounds separate according to their volatility and their interaction with the column coating. In headspace sampling the sample is heated in a sealed vial until its volatile compounds reach equilibrium with the gas above it, and a portion of that gas is injected, so non-volatile matrix does not reach the column. The flame-ionization detector burns the eluting compounds in a hydrogen-air flame, and the ions formed give a current that rises in proportion to the amount of the compound; this signal is used for quantification. The mass spectrometer records a mass spectrum for each peak, which is used for identification.

What controls the result

Headspace response depends on the sample matrix as well as on the compound, so each compound is calibrated with its own reference standard in the same diluent, and the vial temperature and equilibration time are fixed by the method. A compound is identified by its retention time against the standard and by its mass spectrum; a library match without a reference standard is reported as a tentative identification. A blank vial in every batch shows background and carry-over.

ICP-MS

Inductively coupled plasma mass spectrometry

Instrument
Inductively coupled plasma mass spectrometer, after acid digestion of the sample
How it works

A weighed portion is digested in acid, typically nitric acid in a closed vessel under microwave heating, which destroys the organic matrix and retains volatile elements such as mercury. The diluted digest is sprayed into an argon plasma, where the elements are atomized and ionized; the ions are separated by mass-to-charge ratio and counted. Method 4.7 of the FDA Elemental Analysis Manual, for example, reports arsenic at mass 75, cadmium at mass 111, mercury at mass 201 and lead as the sum of masses 206, 207 and 208. Each element is quantified against a multi-element calibration.

What controls the result

Contamination is the main risk at trace level, so method blanks are carried through digestion with every batch. Polyatomic ions that share a mass with an analyte, such as argon chloride on arsenic at mass 75, are reduced in a collision cell, typically filled with helium. Internal standards in every solution correct for drift and for suppression of the signal by the matrix, and a matrix reference material with assigned element values is digested with the batch.

Spectrophotometric and enzymatic assays

Colorimetric and enzymatic assays read by UV-visible spectrophotometry; nitrogen determination for protein

Instrument
UV-visible spectrophotometer; Kjeldahl digestion and distillation or Dumas combustion for nitrogen
How it works

A reagent converts the analyte into a colored product, and the absorbance of the solution at a fixed wavelength is proportional to its concentration (Beer–Lambert law). In the phenol-sulfuric acid assay, carbohydrates form an orange-yellow product that is read at 490 nm against a glucose standard curve. In an enzymatic assay, enzymes of known specificity break down one class of glucan to glucose, and the glucose is measured with glucose oxidase/peroxidase reagent at 510 nm; for mushroom and yeast products, beta-glucan is total glucan minus alpha-glucan. Total protein is the exception in this group: it is calculated from nitrogen, which is determined by Kjeldahl digestion, distillation and titration or by Dumas combustion.

What controls the result

Each batch carries a reagent blank, a standard curve or standard, and a control material of known content, and absorbance is read only within the linear range of the curve. The measurand is defined by the method, so the report states the method, the calibration standard and any conversion factor. For protein, a pure compound of known nitrogen content checks the recovery of nitrogen.

Gravimetric and potentiometric measurements

Measurements by mass, volume and electrode potential

Instrument
Analytical balance, drying oven, muffle furnace, graduated cylinders, pH meter with glass electrode
How it works

Moisture is determined as loss on drying: a weighed portion is dried under fixed conditions and weighed again. Ash is the residue that remains after the organic matter is burnt off in a muffle furnace. Bulk density is the mass of a powder divided by the volume it occupies, untapped, in a graduated cylinder; tapped density uses the volume after tapping. The pH is read with a glass electrode, whose potential changes linearly with pH, and solubility is determined by saturating a solvent at a stated temperature and measuring the dissolved amount.

What controls the result

Each of these results depends on its conditions, so the drying temperature and time, the ignition temperature, the tapping procedure and the measurement temperature are fixed by the method, and the method is stated on the report. Loss on drying counts everything that evaporates under the drying conditions, not only water. Balances are calibrated, and the pH meter is calibrated before use with standard buffer solutions that bracket the pH of the sample.

Culture-based microbiology

Plate count, most probable number, enrichment culture and susceptibility testing of isolates

Instrument
Sterile culture media and diluents; incubators and water baths at the temperatures the methods specify
How it works

For a count, a weighed portion is diluted in tenfold steps, plated on agar and incubated; each viable cell or clump of cells able to grow under those conditions forms one colony, and the result is reported as colony-forming units (CFU) per gram or milliliter. For detection of a specified organism such as Salmonella, a test portion is enriched in broth and streaked on selective agar, and presumptive colonies are confirmed. Coliforms and Escherichia coli are counted on selective media or estimated by the most probable number technique. Antibiotic resistance testing measures the growth of a pure isolate in the presence of each antibiotic, as a zone diameter or a minimum inhibitory concentration.

What controls the result

A count is valid only for the medium, temperature and time of the method, so incubation temperatures are recorded and each lot of medium is tested for growth promotion with reference strains before use. A negative control of diluent and medium without sample must show no growth, and a positive control strain must grow as expected. Before a product type is tested routinely, method suitability shows that the product does not prevent the recovery of a low number of added organisms.

PCR and sequence-based identification

Polymerase chain reaction, agarose gel electrophoresis and marker-gene sequencing

Instrument
Thermal cycler and agarose gel electrophoresis; sequencing of the amplified marker gene
How it works

DNA is extracted from a pure culture, and PCR copies the region between two primers many times over. For identification the region is a taxonomic marker: the 16S rRNA gene for bacteria, the ITS region for fungi and the D1/D2 region of the large-subunit rRNA gene for yeasts. The amplicon is checked on an agarose gel, sequenced on both strands and compared with curated reference sequences. Targeted PCR uses primers specific to one named organism or gene and reports detected or not detected.

What controls the result

PCR amplifies contaminating DNA as readily as target DNA, so every run carries an extraction blank and a no-template control, which must give no product, and a positive control, which must give the expected product. Sequence quality is reviewed before comparison, and the report states the taxonomic level the match supports: species, species group or genus. PCR detects DNA, not living cells, so viability is shown by culture.

Stability storage

Storage at controlled temperature and relative humidity, with testing at fixed time points

Instrument
Storage chambers with controlled temperature and relative humidity
How it works

Units of the product in its final packaging are stored at a defined temperature and relative humidity, and a set is withdrawn at each time point of the protocol and tested for the chosen attributes: active ingredient, physicochemical properties and microbiological quality. ICH Q1A(R2) defines the storage conditions most often used: long term at 25 °C ± 2 °C and 60 % RH ± 5 % RH or at 30 °C ± 2 °C and 65 % RH ± 5 % RH, intermediate at 30 °C ± 2 °C and 65 % RH ± 5 % RH, and accelerated at 40 °C ± 2 °C and 75 % RH ± 5 % RH. Each attribute is reported against time for each storage condition. For microbial formulations the attribute is the viable count.

What controls the result

The storage condition is part of the result, so chamber temperature and humidity are recorded for the whole study and any excursion is stated on the report. Every time point is tested by the same method as the initial test, so that a change reflects the product and not the method. The protocol, with its conditions, time points, attributes and acceptance criteria, is agreed before the study starts.

Controls run with every batch

A sample result is accepted only when every control measured with it meets its criterion. The controls differ by technique. Most run in every batch. Two are tied to something other than the batch and are listed because a batch depends on them: each lot of culture medium is tested before use, and method suitability is shown for each product type before routine testing.

Chemical and biochemical tests

Chemical and biochemical tests
ControlWhat it shows
Method blankAll reagents are carried through the whole preparation without sample, to show contamination from reagents, vessels and the laboratory.
CalibrationReference standards at several concentrations are measured in the same batch, and a result is reported only inside the calibrated range.
System suitabilityBefore any sample, injections of a standard are checked against the criteria of the method for resolution, peak shape and repeatability.
Check standardA standard of known concentration is measured at intervals and at the end of the batch, so that every sample result is bracketed by check standards.
Internal standardWhere the method uses one, an element or compound absent from the sample is added to every solution to correct for drift and matrix effects.
Spiked sampleA known amount of analyte is added to a portion of the sample before preparation, and its recovery shows losses and matrix effects.
Duplicate preparationA second portion of the same sample is prepared and measured separately, which shows the repeatability of the whole procedure on that sample.
Reference or control materialWhere one is available, a material with an assigned or established value is carried through the procedure in the same way as a sample.

Microbiological tests

Microbiological tests
ControlWhat it shows
Negative controlDiluent and medium are carried through the batch without sample and must show no growth.
Positive control strainA reference strain is carried through the same media and incubation and must show the expected growth and reactions.
Growth promotion of mediaEach lot of culture medium is tested with reference strains before use, and selective media are also checked for inhibition of non-target organisms.
Method suitabilityBefore a product type is tested routinely, a low number of reference organisms is added to the prepared sample and must be recovered within the criterion of the reference method.
Replicate platesEach dilution is plated in replicate as the reference method requires, and counts are taken only from plates within its countable range.
Incubation temperature recordIncubator and water-bath temperatures are recorded, because a count is valid only for the temperature and time the method specifies.
Quality-control strain for susceptibility testingA reference strain with published acceptable ranges is tested alongside the isolates, and its zone diameters or MIC values must fall inside those ranges.

Molecular identification

Molecular identification
ControlWhat it shows
Extraction blankExtraction reagents are processed without sample, to show DNA introduced during extraction.
No-template controlA reaction with water in place of DNA is included in every PCR run and must give no product.
Positive controlDNA of a reference strain or of the target organism is amplified in the same run and must give the expected product.
Internal amplification controlWhere the sample can inhibit PCR, a control sequence is amplified in the same reaction, so that a reaction stopped by inhibition is not reported as not detected.

Every result

Every result
ControlWhat it shows
Second-analyst reviewA second analyst checks the raw data, the controls and the calculation before the result is signed.

One determination, step by step

One erinacine A result by UPLC-PDA, from the balance to the signed report. The instrument follows the step being read; select any step to see it again. This is a simulated run with illustrative data, not client data.

Fig. 1Erinacine A by UPLC-PDA, from the balance to the reported result. Simulated run. Illustrative data, not client data.
  1. Sample preparation. We homogenize the sample, weigh 50.00 mg, measure loss on drying (4.0 %), extract into 10.00 mL and dilute 5.00 mL to 20.00 mL (dilution factor 4). Results are reported on a dry-weight basis.
  2. Reference standards. Six calibration levels from 1.00 to 50.00 µg/mL, prepared from an authenticated erinacine A reference standard.
  3. Injection and separation. 1.0 µL onto a C18 column at 40.0 °C and 0.400 mL/min. The gradient runs from 5 % to 95 % acetonitrile over 10 minutes. The detector records a full UV spectrum at every point.
  4. Identification and integration. Retention time within ±0.050 min of the reference standard and a spectrum match of at least 990. Both, or the peak is not identified. Integration is baseline to baseline, with every integration event kept in the audit trail.
  5. External calibration. The sample must fall inside the calibrated range. We never extrapolate: above the top standard we dilute and re-inject.
  6. Calculation and review. 24.00 µg/mL × 4 × 10.00 mL ÷ 48.00 mg = 20.0 mg/g, dry weight. The chromatogram, the integration and the calculation are reviewed before the report is authorized.
Fig. 2System suitability for the erinacine panel by UPLC-PDA. Every criterion is met before a sample result is accepted. Illustrative data.
System suitability, erinacine panel
ParameterCriterionResult
Resolution, erinacine B / A, reference mycelium extract injected with the suitability setAt least 2.02.42
Tailing factor, erinacine AAt most 1.51.06
Plate count, erinacine AAt least 10 00042 400
Area RSD, n = 6, erinacine A standard 10.00 µg/mLAt most 2.0 %0.42 %
Retention-time RSD, n = 6At most 1.0 %0.05 %
Check standard, 10.00 µg/mL95.0 to 105.0 % recovery99.6 %
Fig. 3Injection sequence of one UPLC-PDA batch, in injection order: a blank, six system suitability injections, six calibration levels, a blank, a check standard, four samples in duplicate, a closing check standard and a blank. Every sample result is bracketed by check standards. Illustrative data.
BlankSystem suitabilityCalibrationCheck standardSampleevery sample injection lies between two check standards1.002.505.0010.0025.0050.0010.001122334410.001510152025Injection numberNo.Injection1Blank2System suitability 1 of 63System suitability 2 of 64System suitability 3 of 65System suitability 4 of 66System suitability 5 of 67System suitability 6 of 68Calibration 1.00 µg/mL9Calibration 2.50 µg/mL10Calibration 5.00 µg/mL11Calibration 10.00 µg/mL12Calibration 25.00 µg/mL13Calibration 50.00 µg/mL14Blank15Check standard 10.00 µg/mL16Sample 1, injection 1 of 217Sample 1, injection 2 of 218Sample 2, injection 1 of 219Sample 2, injection 2 of 220Sample 3, injection 1 of 221Sample 3, injection 2 of 222Sample 4, injection 1 of 223Sample 4, injection 2 of 224Check standard 10.00 µg/mL25Blankbetween check standards

Method validation and verification

In-house methods are validated before they are used for reportable results. Chemical methods are validated per ICH Q2(R2), Validation of Analytical Procedures, for specificity, range (response and lower range limit), accuracy and precision. Robustness is tested during method development.

A published standard method is not validated again. It is verified in our laboratory before use: we show that the method performs as it states on the type of sample concerned. USP <467> for residual solvents is an example.

Microbiology methods are shown suitable for each product type. A low number of reference organisms is added to the prepared sample and must be recovered within the criterion of the reference method. If the product inhibits growth, the preparation is changed and checked again.

Methods are transferred with a written protocol and report.

A validation or verification applies to the analyte, the matrix and the range stated in its report. A new matrix or formulation calls for at least a verification.

Acceptance criteria are set for each method in a written protocol before the work starts. The method used, and any modification, is stated on the report.

Data integrity

Audit trails are always on. Every result gets a second-person review. Records are signed electronically.

Observations made by hand, such as weighings, pH readings and colony counts, follow the same principles as instrument data: they are recorded at the time the work is done, and every correction is traceable.

We work to ALCOA+:

ALCOA+
TermMeaning
AttributableEvery entry records who did it and when.
LegibleRecords can be read and understood for as long as they are kept.
ContemporaneousData are recorded at the time the work is done.
OriginalThe first capture of the data is kept, or a verified true copy.
AccurateRecords match what happened, and every correction is traceable.
CompleteAll data are kept, including repeat injections and failed runs.
ConsistentEntries follow the same sequence and time stamps throughout.
EnduringRecords are stored on durable, backed-up media for the retention period.
AvailableRecords can be retrieved for review or audit when they are needed.

Reference basis is the published method a procedure is built on. The method used, and any modification, is stated on the report.